Calibration establishes the true sensitivity of a measurement chain — microphone, cable, input conditioning and analog-to-digital conversion together — by measuring a known reference. Deepsona provides a pistonphone calibration wizard for sound pressure channels, and a separate playback SPL calibration for listening to and analysing recordings.
Pa. The dialog closes.The dialog reads the live level five times per second, so the Live RMS display responds immediately when you fit or remove the calibrator. Values are auto-scaled: volts above 1 V, millivolts above 1 mV, microvolts below that.
Two standard reference levels are offered. Pick the one printed on your calibrator.
| Reference level | Sound pressure | Typical instruments |
|---|---|---|
| 94 dB SPL @ 1 kHz | 1 Pa | Sound level calibrators (e.g. B&K 4231, GRAS 42AB) |
| 114 dB SPL @ 250 Hz | 10 Pa | Pistonphones (e.g. B&K 4220) |
The frequency is informational and does not enter the calculation. What matters is that the sound pressure level produced by your calibrator matches the level you select.
The reference level is converted to a sound pressure in pascals, and the captured RMS voltage is divided by it:
So 94 dB corresponds to 1 Pa and 114 dB to 10 Pa. A microphone that produces 50 mV RMS in a 94 dB calibrator has a sensitivity of 50 mV/Pa.
The result is the sensitivity of the whole chain. If it differs noticeably from the microphone's data-sheet value, the difference is in the cable, the preamplifier or the input conditioning — which is exactly what calibration is meant to capture.
A calibration result lives in the channel row. From there it is preserved in three ways:
| Where | What is kept | How to reuse it |
|---|---|---|
Channel settings file (.vlcs) | Sensitivity, unit and the rest of the channel table. | Save channel settings to a .vlcs file, then import it at the start of the next session with the same frontend. |
| Recording metadata | Sensitivity, sensitivity unit, output unit, sensor identity and calibration date are written into every .dpsa recording. | Automatic. Engineering-unit scaling is applied from this metadata when the recording is analysed. |
| Sensor Library | Nominal and calibrated sensitivity per channel plus a dated calibration history. | Enter the calibrated value in the sensor's Calibrated Sensitivity field and add a Calibration History entry. Assign that sensor to a channel to reuse it. |
A separate calibration governs how an existing recording is interpreted for playback and level analysis. It is set in the toolbar strip labelled SPL Calib:, which defines what a full-scale sample corresponds to in dB SPL.
| Field | Meaning | Default |
|---|---|---|
FS= … dB SPL | The sound pressure level, in dB SPL peak, that a full-scale sample represents. | 114.0 |
Sens= … mV/Pa | Sensor sensitivity. A typical 1/4" microphone is around 50 mV/Pa. Enter 0 if not used. | 0 |
Gain= … dB | Upstream signal-chain gain such as a preamplifier. Enter 0 for none. | 0 |
The summary line next to the fields confirms the current setting, for example
1.0 PCM = 114.0 dB SPL peak · sens 50.0 mV/Pa.
This calibration is stored per source file and is restored automatically the next time you open the same recording, including in later sessions. The same value is used by the audio player, so playback loudness and displayed levels stay consistent.
| Symptom or message | Cause and remedy |
|---|---|
| "No live signal. Start Preview (Online Monitor) first — pistonphone calibration reads live raw V RMS from the running stream." | No data is streaming. Start Preview in the Online Monitor, confirm the frontend is selected and the channel is enabled. |
| Capture stays disabled | No channel is selected, or the live RMS is zero. Check that the channel's On box is ticked and that the calibrator is switched on. |
| Apply stays disabled | Nothing has been captured yet, or the computed sensitivity is zero. Capture a valid reading first. |
| Live RMS reads zero or near zero | The calibrator is off or badly seated, the wrong channel is selected, or the sensor supply is wrong. Check that Supply is set to ICP for a CCLD microphone. |
| Live RMS is unstable | The calibrator is not seated squarely, or there is strong background noise or air movement. Reseat and wait longer before capturing. |
| Sensitivity is about ten times too large or too small | The wrong reference level is selected. Switch between 94 dB and 114 dB — the result recomputes without re-capturing. |
| Sensitivity far from the data-sheet value | Check the input Range (a clipped signal reads low) and the Highpass setting, and confirm you calibrated the channel the microphone is actually plugged into. |
| Levels wrong after re-cabling | The calibration belongs to a specific chain. Re-calibrate after any change to cable, channel, range or supply. |
| Analysis levels differ from live levels | Check the playback SPL Calib: fields for that file — they are stored per file and may carry a value from an earlier session. |
Related: Recorder Setup · Sensor Library and TEDS · Online Monitor