Specific Fluctuation Strength (Hearing Model) resolves the perception of slow amplitude modulation across the frequency axis using the hearing model defined in ECMA-418-2, Edition 4 (2025). Instead of a single overall vacil value, the analysis delivers a distribution of fluctuation strength over 53 half-Bark bands, so that the frequency region responsible for a throbbing or pulsing character can be localized directly.
This analysis complements the classic fluctuation strength model described on the Fluctuation Strength page. The classic model follows the Zwicker/Fastl formulation; the hearing-model variant follows the standardized ECMA-418-2 auditory processing chain. Both are available side by side and are selected according to the reporting requirement of the project.
The input signal passes through the ECMA-418-2 auditory filter bank, which divides the audible range into 53 overlapping bands spaced half a Bark apart, covering 0.5 – 26.5 Bark. Before the filter bank, an ear transfer function corresponding to the selected sound field (free field or diffuse field) is applied. The envelope of every band output is then extracted and analysed for low-rate modulation.
Within each band, modulation components in the fluctuation-relevant range are weighted so that sensitivity peaks near 4 Hz, matching the temporal resolution limit of the auditory system. The band results are produced on a fixed 50 Hz result grid — one value every 20 ms — and a non-linear low-pass stage models the temporal integration (build-up and decay) of the ear, so that short bursts do not create unrealistically sharp peaks in the result.
The band result is a density: fluctuation strength per unit of critical-band rate, expressed in vacil/Bark. Integrating the density over the critical-band-rate axis yields the total fluctuation strength in vacil:
The single value reported in the footer is the 10th percentile F10 of the total fluctuation strength over time — the level that is exceeded during 10% of the evaluated duration. This follows the same percentile convention used elsewhere in psychoacoustics, such as the loudness value N5; it is a statistic of the signal, not a tolerance or an uncertainty. All three variants of this analysis share this same single-value definition, so a band spectrum, a vs Time colour map and a vs RPM colour map computed from the same recording report an identical single value.
| Parameter | Options | Description |
|---|---|---|
| Sound Field | Automatic (default) / Free field / Diffuse field | Ear transfer function of the hearing model. In Automatic mode the sound field is determined per channel from the equalization information stored with the recording; channels without that information fall back to free field |
| Standard Edition | ECMA-418-2 Edition 4 (2025) | Hearing-model edition applied by this analysis |
| Band Resolution | 53 half-Bark bands, 0.5 – 26.5 Bark (fixed) | Band layout defined by the standard; not user-adjustable |
| Result Rate | 50 Hz (fixed) | One result frame every 20 ms; applies to the vs Time variant and to the sampling grid of the vs RPM variant |
| Speed Step | 50 rpm (default) | vs RPM variant only. Spacing of the speed axis |
| Slope | Automatic (default) / Run-up / Coast-down | vs RPM variant only. Direction of the speed change to be evaluated |
| Quantity | Unit | Description |
|---|---|---|
| F′(z) | vacil/Bark | Specific fluctuation strength per band — 53 values, plotted over the band centre frequency |
| F′(z, t) | vacil/Bark | Specific fluctuation strength as a colour map over band and time |
| F′(z, n) | vacil/Bark | Specific fluctuation strength as a colour map over band and speed of rotation |
| F10 | vacil | Single value: 10th percentile of the total fluctuation strength over time; identical for all three variants |
The hearing-model specific fluctuation strength is offered as three analyses that share one computation core and one single-value definition, but differ in how the result is presented:
| Analysis | Result Form | Axes | Typical Use |
|---|---|---|---|
| Specific Fluctuation Strength (Hearing Model) | Band spectrum (one curve per channel) | x = band centre frequency, y = vacil/Bark | Stationary operating points; identifying which frequency region fluctuates |
| Specific Fluctuation Strength (Hearing Model) vs. Time | Colour map | x = time, y = band centre frequency, colour = vacil/Bark | Transient events, load changes, cycle-by-cycle comparison |
| Specific Fluctuation Strength (Hearing Model) vs. RPM | Colour map | x = speed of rotation, y = band centre frequency, colour = vacil/Bark | Run-up and coast-down; locating speed ranges with pronounced fluctuation |
The hearing model requires an ear transfer function, and the correct choice depends on how the signal was recorded. The analysis therefore offers an Automatic setting that reads the equalization information stored with each recorded channel: